Volatiles Analysis Using Static Headspace Sampling

  1. 1,1-Dichloroethene
  2. trans-1,2-Dichloroethene
  3. 1,1-Dichloroethane
  4. cis-1,2-Dichloroethene
  5. Bromochloromethane
  6. Chloroform
  7. 2,2-Dichloropropane
  8. 1,2-Dichloroethane
  9. 1,1,1-Trichloroethane
  10. 1,1-Dichloropropene
  11. Carbon tetrachloride
  12. Dibromomethane
  13. 1,2-Dichloropropane
  14. Trichloroethene
  15. Bromodichloromethane
  16. cis-1,3-Dichloropropene
  17. trans-1,3-Dichloropropene
  18. 1,1,2-Trichloroethane
  19. 1,3-Dichloropropane
  20. Dibromochloromethane
  21. 1,2-Dibromomethane
  22. Tetrachloroethene
  23. 1,1,1,2-Tetrachloroethane
  24. Bromoform
  25. 1,1,2,2-Tetrachloroethane
  26. 1,2,3-Trichloropropane
  27. Bromobenzene
  28. 1,3-Dichlorobenzene
  29. 1,4-Dichlorobenzene
  30. 1,2-Dichlorobenzene
  31. 1,2-Dibromo-3-chloropropane
  32. 1,2,4-Trichlorobenzene
  33. Hexachlorobutadiene
  34. 1,2,3-Trichlorobenzene

Column: DB-VRX, 60m x 0.32mm I.D., 1.8µm
Order No.: 123-1564
Carrier: Helium at 50cm/sec
Oven: 35°C for 8 min, 35-220°C at 10°/min, 220°C for 4 min
Injector: Split 1:20, 200°C
Detector: ECD, 300°C nitrogen makeup gas at 90mL/min, baseline corrected

Headspace Analyzer Conditions, Tekmar 7000
Vial volume: 22.5mL, Water volume: 16.5mL
Platen temperature: 75°C, Sample equilibration time: 25.0min
Loop fill time: 0.03min, Loop equilibration time: 0.1min
Loop size: 1.0mL, Injection time: 3.0min

 GC Chromatograms

HOME

                                                           Vial Quote  | Logistic Services  | Policies / TermsInternational Terms
________________________________________________________________________________________________________
              All prices subject to change without notice.

*PRICING NOTE* - P.J. Cobert Associates, Inc. would like to notify you that we have received late notification of price increases

for several manufacturers effective August 1, 2008. We have no control over when manufacturers increase their prices, however

we always do everything possible to keep our pricing as competitive as we can. We apologize for any inconvenience this may cause.

________________________________________________________________________________________________________

Send mail to cobert@cobertassoc.com with questions or comments about this web site.
Copyright © 2005-2008 COBERT ASSOCIATES    All rights reserved
No part of this website may be reproduced or transmitted in any form or by any means, mechanical or electronic, including photocopying or recording or by any information storage and retrieval system, without written permission from an officer of P.J. Cobert Associates, Inc.